Gamry Instruments EIS Box Скачать руководство пользователя страница 19

General Operation  

19 

 

 

 
The 

system computer

 also controls the 

instrument

. When a cell is connected to the 

instrument

 cell leads, the 

computer causes a test to be run on that cell. The test can be as simple as a single current or potential 
measurement or as complex as a cyclic scan. The computer software and 

instrument

 are responsible for taking 

the measurement and storing it.   
 
The EIS Box 

instrument

 in Figure 3-

1 is shown as a “black box”. The relays in the 

EIS Box can carry currents up 

to 5 A. 
 
In a typical EIS Box experiment, the 

system computer

 cycles through the cells. Each cell is connected in turn to 

the 

instrument

 and used to collect a measurement. Usually the computer delays at the end of each cycle 

through the cells so that the measurements on each cell are separated by a fixed time period. 

Keyword Definitions 

Active Cell 

The active cell is the cell (if any) currently connected to the system potentiostat. Any electrochemical 
measurements are made on the active cell. 
 

Cell 

The term cell is used to describe either an electrochemical cell or the switches needed to connect an 
electrochemical cell to the instrument.   
 

Channel

 

An EIS Box contains eight channels. A channel includes cell-switching relays. The term cell is used 
when only the cell-switching portion of a channel is being discussed. 
 

Inactive Cell 

The term inactive cell is any cell not currently connected to the instrument. 
 

Instrument 

This is the instrument responsible for making electrochemical measurements on the cells used in the 
experiment. The instrument is typically an EIS Box. 
 

System Computer 

The computer responsible for coordinating operation of an EIS Box-based electrochemical test system, 
not to be confused with the very simple micro-controller buried within the EIS Box. 

Switching and Inactive Cell Control 

In the Operation Overview, we said you can think of the EIS Box as a complex switch with controller. Each of 
the eight EIS Box channels is made up of an independent set of switches and control circuits. All the switches 
are under control of the system computer. The system computer must insure that only one cell at a time is 
connected to the instrument, therefore a main switch controls which cell is active. 

 

 

 

Communications Overview 

The system computer communicates with the EIS Box via an ethernet connection or a USB cable.  

 Switching from channel to channel involves electromechanical relays. You will hear a click 

each time relays activate during the switching process. 

Содержание EIS Box

Страница 1: ...EIS Box Instrument Operator s Manual Copyright 2018 Gamry Instruments Inc Revision 1 0 November 1 2018 Gamry P N 988 00071...

Страница 2: ...2...

Страница 3: ...in installation or use of a system containing an EIS Box please try to call from a phone next to your computer where you can type and read the screen while talking to us We will be happy to provide a...

Страница 4: ...ny time without incurring any obligation to install same on systems previously purchased All system specifications are subject to change without notice There are no warranties which extend beyond the...

Страница 5: ...truments Inc assumes no responsibility for errors that might appear Copyrights EIS Box Instrument Operator s Manual copyright 2018 Gamry Instruments Inc all rights reserved Gamry Framework copyright 1...

Страница 6: ...6...

Страница 7: ...Cell Control 19 Communications Overview 19 Chapter 4 Installation 21 Initial Visual Inspection 21 Physical Location 21 Computer Requirements 22 Quick Start Guide for System Installation 22 Software I...

Страница 8: ...erential Electrometer 43 Voltage Measurement 43 Current to Voltage Converter 44 Current Measurement 44 Environmental 44 General 44 Channel Characteristics 45 Appendix B EIS Box Cell Connectors 47 Appe...

Страница 9: ...cuitry does not contain voltages higher than 15 V DC Generally input and output voltages in the EIS Box are limited to 9 V This voltage level is considered safe The EIS Box is normally provided with a...

Страница 10: ...y and other surge events such as lightning Warning Never connect the Protective Ground and System Ground if the electrochemical system you are testing is earth grounded Two earth ground connections in...

Страница 11: ...l Stresses Treat your EIS Box as potentially hazardous if any of the following is true of the unit It shows visible damage It does not operate properly It has been stored for a long period of time und...

Страница 12: ...se isopropyl alcohol Do not use a wet rag or allow fluid to enter the EIS Box enclosure Do not immerse the EIS Box in any type of cleaning fluid including water Do not use any abrasive cleaners Servic...

Страница 13: ...ectrical equipment for measurement control and laboratory use EMC Requirements describing acceptable limits for electrical transient susceptibility in laboratory test equipment It should continue to o...

Страница 14: ...Safety Considerations 14...

Страница 15: ...esigned to for battery testing via electrochemical impedance spectroscopy up to eight cells The EIS Box excels at electrochemical impedance testing of small to medium sized single cell energy storage...

Страница 16: ...t software Software and Applications The EIS Box supports Gamry Instruments electrochemical impedance spectroscopy software under the Framework system Typical applications for the EIS Box include Rese...

Страница 17: ...ts from other general purpose switching devices The EIS Box is integrated into all Gamry Instruments electrochemical impedance spectroscopy applications that run under the Gamry Instruments Framework...

Страница 18: ...ritical component of the system It controls the EIS Box by sending it simple commands through an ethernet or USB connection The most important function of these commands is to select which cell is act...

Страница 19: ...l An EIS Box contains eight channels A channel includes cell switching relays The term cell is used when only the cell switching portion of a channel is being discussed Inactive Cell The term inactive...

Страница 20: ...20...

Страница 21: ...o the rear of the instrument because some cable connections are made from the rear The EIS Box is generally operated in a flat position see Figure 4 1 Operation in other orientations is possible as lo...

Страница 22: ...ation so the information provided below is always subject to change Software Installation The EIS Box is compatible with the Windows Plug Play configuration system Like most Plug Play hardware it is b...

Страница 23: ...r of the EIS Box Fig 4 2 The power supply provided inside the EIS Box is rated for operation from 100 to 240 V AC at frequencies from 47 to 63 Hz It should therefore be useable worldwide The EIS Box i...

Страница 24: ...status of the other LED indicators is not important at this time Ethernet Cables The EIS Box connects to the computer using a shielded CAT5E Ethernet cable see figure below The Ethernet connection sho...

Страница 25: ...USB cable Cell Cable Installation The Cell Cables connect pairwise to two 15 pin D type connectors located on the front of the EIS Box see Figure 4 1 for each channel There are two cables per channel...

Страница 26: ...brate instruments connected to the system Framework Device Status Bar By default the Gamry Framework shows a Device Status Bar under its main menu see Figure 4 3 If you do not see the Device Status Ba...

Страница 27: ...dles most of the functions of the EIS Box The second is the Communications Firmware This program handles the communications between your EIS Box and the host computer The third is the PLD firmware whi...

Страница 28: ......

Страница 29: ...em A calibration utility is provided with the Gamry Framework The calibration for the EIS Box is divided into two sections Instrument DC Calibration and cable calibration Gain access to these calibrat...

Страница 30: ...been serviced You notice breaks or discontinuities in the data curves recorded with your system The system is being run in an environment that is very different from the previous operating environmen...

Страница 31: ...ery EIS Box is shipped with eight pairs of standard shielded cell cables The Gamry part number for the Current Cable is 985 00183 and the Sense Cable is 985 00184 They are both 3 m complex cables with...

Страница 32: ...power amplifier If your cell is a typical glass laboratory cell all of the electrodes are isolated from earth ground In this case you may be able to reduce noise in your data by connecting the EIS Box...

Страница 33: ...ence the term battery is used here to refer to all single cell or stacked electrochemical devices that can distribute energy Grounding errors on a battery can be particularly dangerous for they can re...

Страница 34: ...resistance of a good fuse is very small Figure 6 3 Drawing of Fuse holder in the D Connector Hood Four replacement fuses should accompany every Counter Working cell cable shipped by Gamry Instruments...

Страница 35: ...Utilities Set a Voltage on the Framework software s menu bar You see a Set a Voltage dialog box similar to this Figure 6 4 Setup Dialog Box for the Set a Voltage Script In the Voltage field enter a Vo...

Страница 36: ...current readings are near zero and a red CA Overload indication may be seen at the bottom of the runner window If this test indicates an open fuse use the procedures described above to check both fus...

Страница 37: ...ff There is no AC mains supply connected to the rear panel Power In connector One of the internal DC power supplies has failed Cell Connectors The EIS Box has two sets of eight cell cable connectors f...

Страница 38: ...el LED is between each pair of Cell connectors The Channel LEDs glow amber whenever the EIS Box is actively applying voltage or current to the electrochemical cell attached to that channel Avoid touch...

Страница 39: ...you can connect to the Chassis Ground jack Use a wire to connect the banana plug to earth ground A black banana plug to banana plug lead is provided with your EIS Box You may find it useful when makin...

Страница 40: ...ess point to the protective earth ground in the EIS Box It is connected to the third wire in the AC power cord and to the EIS Box s metal chassis This binding post can be connected to an additional ea...

Страница 41: ...on the left side of the EIS Box s rear panel It can be identified by the line cord connector built into it 3 Find the small slot on the right side of the AC input module near the line voltage setting...

Страница 42: ...Make sure you have not inadvertently changed the AC line voltage setting which is displayed in a small window on the AC input module If the voltage shown is not the appropriate setting for the line v...

Страница 43: ...Amplifier Compliance Voltage Min 6 5 V Note 1 Output Current Min 5 A Note 2 Unity Gain Bandwidth Typ 1050 250 43 4 4 0 5 kHz Note 3 Slew Rate Typ 10 4 4 0 7 0 07 0 007 V s Note 4 Differential Electrom...

Страница 44: ...5 MHz Note 17 Current Measurement Resolution Typ 0 00333 FS bit Offset Range Typ 100 of range Post offset Gain Typ 1 10 100 Accuracy Typ Dominated by current to voltage error see above Environmental O...

Страница 45: ...2 G resistor switched into the input and measuring the voltage difference with and without the input resistance 7 The differential impedance is measured between the Reference and Work Sense inputs Th...

Страница 46: ...The specified I E Converter BW is measured at 5 mA I E Range 18 The total error in a voltage setting is Error DC Zero Offset Gain Voltage Setting For a 1 V signal the theoretical error can be as high...

Страница 47: ...Ground The potentiostat s floating ground Can be used to shield the cell if very low currents need to be measured Also used as a shield for the counter electrode cable 5 12 Counter Connected to the c...

Страница 48: ...the cell end of the cell cable 7 15 Work Sense Shield The shield for the work sense electrode input Driven to the same potential as Pin 8 Left open at the end of the cell cable 8 Work Sense Connected...

Страница 49: ...Appendix C CE Certificate 49 Appendix C CE Certificate Certificate of Conformance...

Страница 50: ...50...

Страница 51: ...fan speed 12 firmware update 29 Floating Ground binding post 11 41 42 Framework software 17 28 31 fuses 35 36 37 38 43 Gamry Framework 17 Gamry Instrument Manager 28 31 green cell lead 34 ground 49 50...

Страница 52: ...52 Work Shield 49 Working electrode 33 34 35 39 49...