Installation
24
Power-up Test
Before you make any other connections to your EIS Box, check that the EIS Box is at least nominally functional.
After connecting AC power to the EIS Box, toggle the
Power
switch on the front panel of the EIS Box.
Watch the
Power
LED as the EIS Box powers up. It glows a steady blue. If you do not see a continuous blue
color, contact technical support.
The status of the other LED indicators is not important at this time.
Ethernet Cables
The EIS Box connects to the computer using a shielded CAT5E Ethernet cable (see figure below).
The
Ethernet
connection should not
be “hot
-
plugged”.
This means both the computer and the EIS Box must be
powered OFF before the ethernet cable is plugged in.
Caution:
If the
Power
LED goes on, then turns off and stays off, the EIS Box is not
working properly! If this power-up test fails, contact Gamry Instruments or your local Gamry Instruments
representative as soon as possible.
Power
LED
Power
switch
Ethernet