WRO 300 H
HCEN12104 Rev 09.2010
Technical data
47
4.4 Environmental data
4.4.1
Electromagnetic environment
Value
Operation
Ambient Temperature range
+ 10 to + 40 °C
Relative Humidity range
30 to 85 % RH
Air Pressure range
700 to 1060 hPa
Transportation
and storage
Ambient Temperature range
- 10 to + 40 °C
Relative Humidity range
10 to 95% RH
Air Pressure range
500 to 1060 hPa
For transportation and storage below the freezing point,
the WRO 300 H must be filled with a preservation solu-
tion of 2 % MEMSTOR, at least 9% glycerol and the
remaining parts RO-water.
During transportation and storage the equipment must be
kept in its original packing. If transportation or storage
time is more than 15 weeks, the environmental data relat-
ing to the operation must be followed.
If condensation occurs when moving the equipment
between locations with different temperatures and high
relative humidity (e.g. outdoor and indoor locations), the
inside of the equipment must be allowed to dry before
switching on the equipment.
Value
The WRO 300 H is intended for use in the electromagnetic environment specified below.
The customer or the user of the WRO 300 H should assure that it is used in such an
environment.
Emissions
test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group1
The WRO 300 H uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment
Содержание WRO 300
Страница 8: ...WRO 300 H 8 HCEN12104 Rev 09 2010...
Страница 38: ...WRO 300 H 38 Regular maintenance HCEN12104 Rev 09 2010...
Страница 64: ...WRO 300 H 64 Troubleshooting HCEN12104 Rev 09 2010...