WRO 300 H
HCEN12104 Rev 09.2010
Troubleshooting
53
5 Troubleshooting
In alarm situations the message text will appear in the display.
Only the latest message will be visible in the display. By using
the
,
or
the display will be unlocked and the alarm
list can be scrolled. Main view will reappear after a time period.
5.1 Alarms and notifications
5.1.1
General
No.
Alarm and
notifications
Definition
101
CONDUCTIVITY
SENSOR FAIL-
URE
WRO 300 H Buzzer sounds with Alarm sound,
flashes, message text is displayed.
Reason
The initial conductivity system check failed.
Action
1
Press
,
or
to silence
buzzer
2
Check that the product water conductivity
is within normal range. If so, continue as
normal but notify technical personnel.
102
HIGH LEVEL
SENSOR FAIL-
URE
WRO 300 H WRO 300 H stops, buzzer sounds with Stop
sound,
flashes,
message text is displayed.
Reason
Failure of High level sensor in tank.
Action
1
Press
,
or
to silence
buzzer
2
Notify technical personnel.