WRO 300 H
52
Technical data
HCEN12104 Rev 09.2010
4.5 Safety
d. Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electro-
magnetic site survey should be considered. If the measured field strength in the
location in which the WRO 300 H is used exceeds the applicable RF compliance
level above, the WRO 300 H should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the WRO 300 H.
e. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3 V/m.
f.
NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
The machine complies with the following standards:
IEC 60601-1 General requirements for safety, Class I, type B, and IPX1.
IEC 60601-1-2 Electromagnetic compatibility.
Содержание WRO 300
Страница 8: ...WRO 300 H 8 HCEN12104 Rev 09 2010...
Страница 38: ...WRO 300 H 38 Regular maintenance HCEN12104 Rev 09 2010...
Страница 64: ...WRO 300 H 64 Troubleshooting HCEN12104 Rev 09 2010...