WRO 300 H
HCEN12104 Rev 09.2010
Regular maintenance
29
9
Initiate the disinfectant intake by pressing
until light
goes on.
Pause, interrupt or restart procedure
10
401 REMOVE WAND
is displayed and the buzzer sounds
and
is flashing, press
.
NOTE
The display shows present phase of the Chemical disinfection proce-
dure.
NOTE
The container can be empty before the intake phase is finished.
CAUTION
The Chemical disinfection has to be restarted if alarm
406 INSUF-
FICIENT INTAKE
has appeared.
Action
Pause of chemical
intake phase
Press
until light flashes. To continue the
intake press
again until light goes on.
Interrupt Chemical
disinfection procedure
Disconnect the wand connector and press
until light goes on, to initiate rinse.
Restart of intake phase
Remove the wand connector and insert it again
CHEM INTAKE RESTART
appears on the display
and
flashes. Then press
until light
goes on .
CHEM INTAKE
X
μ
S
/
cm