Regular maintenance
HCEN12104 Rev 09.2010
3.2.4
Central Chem
The Central Chem disinfects the WRO and also sends
disinfectant into the dialysis machine.
For more detailed information on how to perform Central Chem
together with Gambro dialysis machines, please refer to
separate instructions for Central Chem
.
Procedure
Follow the
“Chemical disinfection procedure” on page 27
, but
start at
step 3
to and follow the instruction until
step 13
.
In
step 8
select the preferred Central Chem protocol in the
CHEM SELECT
menu.
CAUTION
If the produced disinfection solution is fed into a device other than
Gambro dialysis machines, it is the user’s responsibility to ensure
that the connected device is compatible with Peracetic acid
disinfectants and to establish the procedure for the WRO in
combination with the connected device.
CAUTION
As the disinfection solution produced by the WRO will enter any
device connected to the quick connector on the product water loop,
residual test after disinfection must be performed on the connected
device in accordance with the Operator’s manual for the connected
device.
NOTE
During the intake phase of the Central Chem program, the product
water loop will not be pressurized until the correct conductivity has
been obtained, which will take a few minutes.
CENTR_CH_95
CENTR_CH_200
CENTR_CH_200U