WRO 300 H
48
Technical data
HCEN12104 Rev 09.2010
RF emissions
CISPR 11
Class B
The WRO 300 H is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes
Harmonic
emissions IEC
61000-3-2
Class A (Not
applicable for
115 V version)
Voltage fluctu-
ations/flicker
emissions IEC
61000-3-3
Complies (Not
applicable for
115 V version)
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment-
guidance
Electrostatic
discharge
(ESD)IEC
61000-4-2
±6 kV contact
±8 kV Air
±6 kV contact
±8 kV Air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Surge IEC
61000-4-5
±1 kV for dif-
ferential mode
±2 kV for com-
mon mode
1 kV for differ-
ential mode
± 2kV for com-
mon mode
Mains power quality should be that
of a typical commercial or hospital
environment
Voltage dips,
short interrup-
tions and volt-
age variations
on power sup-
ply input lines.
IEC 61000-4-
11
<5% U
T
a
(>95% dip in
U
T
) for 0.5
cycle
<5% U
T
(>95%
dip in U
T
) for
0.5 cycle
Mains power quality should be that
of a typical commercial or hospital
environment.
40% U
T
(60%
dip in U
T
) for 5
cycles
40% U
T
(60%
dip in U
T
) for 5
cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
<5% U
T
(>95%
dip in U
T
) for 5
sec
<5% U
T
(>95%
dip in U
T
) for 5
sec
Value