Revision 5
92
August 2021
Many of the parameters on this page are also available for editing on the analyzer’s local
display using the analyzer keypad, though not all of them are. Table 24 gives an overview of
the parameters available for editing on this page.
Table 24: Configuration Parameters
Parameter
Editable
via
Keypad
Local Display
Editing Location
Explanation
Cell Temperature Set
Point
Y
Config – Timers/Alarms Control setpoint for analyzer’s measurement
cell heaters.
Low Cell Temperature
Alarm
Y
Config – Timers/Alarms Cell temperature below which a cell
temperature fault will be triggered
High Cell Temperature
Alarm
N
N/A
Cell temperature above which a cell
temperature fault will be triggered
Cell Temperature
Proportional Band
Y
Config – Timers/Alarms Cell temperature PID control parameter
Cell Temperature Reset
Time
N
N/A
Cell temperature PID control parameter
Probe Temperature Set
Point
Y
Config – Timers/Alarms
Control setpoint for probe cooler solenoid
actuation
NOTE: Only if probe control is set to RTD
Probe Temperature
Dead Band
Y
Config – Timers/Alarms
Number of degrees below probe
temperature setpoint measured probe gas
temperature must drop prior to probe cooler
solenoid turning off
NOTE: Only if probe control is set to RTD
Low Probe Temperature
Alarm
Y
Config – Timers/Alarms Probe gas temperature below which low
probe temperature fault will be triggered
High Probe Temperature
Alarm
Y
Config – Timers/Alarms Probe gas temperature above which high
probe temperature fault will be triggered
Cabinet Temperature Set
Point
Y
Config – Timers/Alarms Control cabinet temperature above which
the cabinet cooler will be turned on
Cabinet Temperature
Dead Band
Y
Config – Timers/Alarms
Number of degrees below cabinet
temperature setpoint the cabinet
temperature must drop prior to the cabinet
cooler switching off
Low Cabinet
Temperature Alarm
N
N/A
Control cabinet temperature below which a
cabinet temperature warning will be
triggered
High Cabinet
Temperature Alarm
N
N/A
Control cabinet temperature above which a
cabinet temperature warning will be
triggered.
High SVAP Warning
Y
Config – Timers/Alarms Sulfur vapour concentration above which a
High Sulfur Vapour warning will be triggered
High SVAP Alarm
Y
Config – Timers/Alarms
Sulfur vapour concentration above which a
high sulfur vapour fault will be triggered
SVAP Set Point
N
N/A
Sulfur vapour concentration above which the
probe cooler will be turned on
NOTE: Only if probe control is set to
SVAP
SVAP Dead Band
N
N/A
Amount the sulfur vapour concentration
must drop below the set point before the
probe cooler switches off
NOTE: Only if probe control is set to
SVAP
Filter Points
N
N/A
Number of data points averaged together to
produce the displayed concentration values
Auto Cal Intervall (min)
Y
Config – Timers/Alarms
Time interval from last zero calibration cycle
(manual or automatic) to next scheduled
auto zero calibration cycle.
Содержание 943-TGS-CE
Страница 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Страница 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Страница 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Страница 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Страница 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Страница 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Страница 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Страница 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...