Figure 48: Analysis Section - Analysis Page
The
Analysis
page is divided into 5 sections, each displaying different information.
5.2.1.1
Value Display
The
Value Display
section displays concentration data, calculated analysis data, and
analyzer parameter as described in Table 17.
Table 17: Value Display Parameters
Parameter
Description
Air Demand
Present air demand, in % excess air based on currently measured H
2
S and SO
2
concentrations, calculated as per equation in Section 1.1.
H2S
Present calculated concentration of H
2
S in percent
SO2
Present calculated concentration of SO
2
in percent
COS
Present calculated concentration of COS in percent
CS2
Present calculated concentration of CS
2
in percent
NOTE: CS
2
is an optional measurement. Only analyzers configured with this
option will be capable of measuring CS
2
Sulfur Vapour
Present calculated concentration of sulfur vapour in percent
Cell Pressure
Present measured pressure inside the sample measurement cell in mmHg
Cell Temperature
Present measured concentration of the measurement cell block assembly in °C
Probe Temperature
Present measured temperature of the sample gas exiting the probe in °C
Cabinet Temperature
The present measured temperature of the interior of the analyzer’s control cabinet
in °C.
Содержание 943-TGS-CE
Страница 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Страница 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Страница 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Страница 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Страница 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Страница 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Страница 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Страница 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...