Revision 5
70
August 2021
4.8.2.3
Fixed Temperature and Pressure
The calculated concentrations are compensated for both temperature and pressure.
Typically, this temperature and pressure compensation uses the live values from the
analyzer’s cell temperature and pressure measurement sensors. However, there is also
the capability to use static values of temperature and pressure for this compensation. To
use a
Fixed Temperature
and / or a
Fixed Pressure
value for compensation of
concentration values, use the F4 / F3 keys to highlight the checkbox of the fixed value to
be used, then press
↵
to place an X in the box.
4.8.2.4
Sample and Zero Sample Rate
The values shown in the
Sample Rate
and
Zero Sample Rate
fields determine how
frequently the analyzer updates the displayed concentrations and air demand when the
analyzer is in sampling mode (Sample Rate) and in back purge mode (Zero Sample rate).
The time values entered in these two fields are in units of milliseconds. The default value
for each sampling rate is 500 ms.
4.8.3 Display
The
Display
sub-panel, shown in Figure 42, allows the user to configure certain parameters
that affect how data is displayed on the analyzer’s local display screen.
Figure 42: Display Sub-Panel
4.8.3.1
Analysis 1 and 2 Trend Graph Ranges
The displayed x and y axes for the trend graphs shown on the Analysis 1 and Analysis 2
panels can be configured by the user on this panel using the
Analysis 1/2 X Range
and
Analysis 1/2 Y Range
data entry fields. The Y ranges are concentrations, affecting the
displayed vertical ranges on the Air Demand trend on the
Analysis 1
panel and the
Component concentration trends on the
Analysis 2
panel. These scales can be adjusted
if any of the trend lines are found to be commonly going off of the currently displayed
scale. The X ranges are time, affecting the amount of time represented by the displayed
Содержание 943-TGS-CE
Страница 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Страница 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Страница 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Страница 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Страница 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Страница 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Страница 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Страница 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...