Revision 5
129
August 2021
Steam Purging the Sample Probe
is to be performed by trained and authorized
personnel only.
To perform the a steam purge of the sample probe assembly, follow the procedure below: :
1. Place the analyzer in Offline mode by toggling the
Online / Offline
control toggle
(Section 4.3.1) on the
Analysis 2
panel to
Offline
(red).
2. Place the analyzer in back purge mode by pressing the F6 key on the keypad OR by
placing the X in the
Back Purge
check box on the
Analysis 1
panel.
3. Open the oven cabinet door and remove the black lid on the oven enclosure.
The control cabinet door may NOT be opened while the analyzer is
energized UNLESS the area is known to be non-hazardous. Observe all the
warning labels on the analyzer enclosures.To open the control cabinet door
while the analyzer is energized, the purge control box must first be
bypassed. Refer to Section 3.3.
The oven enclosure and all of its components are HOT (150ºC/300ºF). Wear
appropriate personal protective equipment (gloves, eye wear, clothing etc)
if you are working with hot surfaces or wait until surface have cooled to safe
temperatures before performing maintenance tasks.
4. Close valves V1, V3, and V5. Prior to performing a steam purge, Galvanic Applied
Sciences also recommends disconnecting the measurement cell side fittings on valves
V1 and V3 and install plugs on these valves. This will prevent steam or condensate
from entering the measurement cell.
5. Blow down the steam supply lines to be used for steam purging the sample probe
assembly prior to connecting to the analyzer to clear the line of any condensate.
6. Connect steam supply lines to the Sample Side Access and/or Vent Side Access ports
on the left side of the analyzer cabinet.
7. Switch valves V2 and V4 to the ON position. This will begin the steam purge process.
Zero gas will be venting into the oven when the tubing is removed.
8. Re-install the oven enclosure cover. Ensure that the fibre optic cables are still routed
in the indent in the oven enclosure frame to avoid damage when the oven enclosure
cover is installed.
9. Steam Purge the Sample Probe for between 15 and 30 minutes.
10. When purging is complete, remove the oven enclosure cover.
11. Switch valves V2 and V4 to the OFF position.
12. Disconnect the steam supply lines from the Probe Access ports on the left side of the
analyzer cabinet. Replace any sections of tubing that connect the measurement cell to
Содержание 943-TGS-CE
Страница 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Страница 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Страница 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Страница 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Страница 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Страница 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Страница 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Страница 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...