Precautions about Magnetic Resonance Imaging (MRI)
WARNING
Do not use this equipment in magnetic resonance imaging (MRI)
environments.
When conducting MRI test, remove the electrodes and sensors
connected to the patient. The local heating caused by the induced
electromotive force may cause burn injury to the patient. For details,
refer to the operation manual for the MRI testing device.
Electrosurgery Safety
WARNING
The monitoring system contains protection against interference generated
by electrosurgical instruments. However, depending on the operating
conditions, surgery site with respect to the location of ECG electrodes,
ground plate attachment condition, or the type of instrument used, it may
cause burn injury at the electrode site or noise on the ECG. The noise is
generated at the tip of the electrosurgical knife and is difficult to completely
eliminate because of the frequency components of the ECG. To reduce
electrosurgical interference, take the following precautions:
Location:
Locate the electrosurgical unit as far as possible from this equipment and
the patient cable. This will help reduce interference on the ECG through
the monitor or cables.
Electrode Placement:
The amount of noise interference is considerably different depending on
the electrode position and surgery site. Place the ECG electrodes as far
away as possible from the surgery site and the ground plate. Do not
place electrodes in the path between the surgery site and the ground
plate. If the electrodes are placed in this path, the amount of interference
will be quite large. Position (+) and (–) electrodes as close as possible to
each other.
Ground Plate:
When using electrosurgical instruments, make sure the contact between
the patient and the ground plate is secure. If the connection is
incomplete, the patient may suffer from burn at the electrode site.
Содержание LX-8000 Series
Страница 1: ......
Страница 2: ......
Страница 22: ...xx ...
Страница 26: ...1 General Description 2 ...
Страница 34: ...3 Preparation 10 ...
Страница 40: ...5 Respiration Monitoring 16 ...
Страница 54: ...6 SpO2 Monitoring 30 ...
Страница 74: ...9 Other Setting Items 50 ...
Страница 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Страница 84: ...11 Troubleshooting 60 ...
Страница 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Страница 102: ...14 Standard and Optional Accessories 78 ...
Страница 108: ...15 Specification 84 ...
Страница 109: ......
Страница 110: ......