6. SpO
2
Monitoring
26
●
Applying the LNCS inf-L, LNCS inf-3, RD SET Inf
1. Open the pouch and remove the sensor. Remove the backing from the
sensor, if present.
2. Direct the sensor cable so that it either points away from the patient or
runs along the bottom of the foot. Position the detector onto the fleshy
part of the great toe (Fig 1). Complete coverage of the detector window is
needed to ensure accurate data.
3. Wrap the adhesive wrap around the toe and ensure that the emitter
window (red star) aligns on the top of the toe directly opposite the detector
(Fig. 2).
Verify correct positioning and reposition if necessary (Fig. 3).
1
2
3
●
Applying the LNCS Neo-L, LNCS NeoPt-L, LNCS NeoPt3,
RD SET Neo, RD SET NeoPt
1. Open the pouch and remove the sensor. Remove the backing from the
sensor, if present.
2. For fragile skin, the stickiness of the medical grade adhesive can be
diminished or eliminated by daubing the adhesive areas with a cotton ball
or with gauze (Fig. 1).This step does not apply to the NeoPt 500.
3. Direct the sensor cable so that it either points away from the patient or
runs along the bottom of the foot. Apply the detector onto the fleshy part
of the lateral aspect of the sole of the foot aligned with the fourth toe (Fig.
2). Alternatively, the detector may be applied to the top of the foot (not
shown). Complete coverage of the detector window is needed to ensure
accurate data.
4. Wrap the adhesive/foam wrap around the foot and ensure that the emitter
window (red star) aligns directly opposite of the detector. Be careful to
maintain proper alignment of the detector and emitter windows while
attaching adhesive/foam wrap to secure the sensor (Fig. 3).
Verify correct positioning and reposition if necessary. (Fig. 4).
Содержание LX-8000 Series
Страница 1: ......
Страница 2: ......
Страница 22: ...xx ...
Страница 26: ...1 General Description 2 ...
Страница 34: ...3 Preparation 10 ...
Страница 40: ...5 Respiration Monitoring 16 ...
Страница 54: ...6 SpO2 Monitoring 30 ...
Страница 74: ...9 Other Setting Items 50 ...
Страница 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Страница 84: ...11 Troubleshooting 60 ...
Страница 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Страница 102: ...14 Standard and Optional Accessories 78 ...
Страница 108: ...15 Specification 84 ...
Страница 109: ......
Страница 110: ......