vi
CAUTION
5. After using the equipment, verify following items.
Return all operating switches, knobs etc to the position before using the
equipment, and then switch off the power.
When unplugging the cables, make sure to pull from the connector part of
the cable to avoid excessive force on the cable.
Clean the accessories and cables, and keep them together in one place.
Keep the equipment clean to ensure proper operation for the next use.
Make sure to remove the batteries if the equipment is not used for a long
time. The leakage from the batteries may damage the equipment, or an
explosion from the batteries may occur.
6. If the equipment is damaged, do not attempt service. Ensure patient
safety by immediately turning off the power and removing the
electrodes and cables from the patient. Label the unit “OUT OF
ORDER” and contact your nearest service representative.
7. Do not disassemble or remodel the equipment.
8. Maintenance and Inspection
Make sure to periodically check the equipment and parts. (It is
recommended to conclude a maintenance contract.)
Before reusing the equipment that has been left unused for a while, make
sure that the equipment operates properly and safely.
9. When using electrosurgical knives or defibrillator with this equipment,
follow the precautions below.
To prevent patient from burn injury, verify proper attachment of patient
ground plate, ECG electrode type when using the electrosurgical knife, and
verify paste volume, output energy when using the defibrillator.
Some types of equipment other than the above may cause accidental
hazards to the patient and operator due to the conditions of the equipment.
Read the operation manual attached to each equipment and fully
understand the precautions prior to use.
Non-Explosion Proof
DANGER
Never operate the equipment in the presence of flammable anesthetics,
high concentration of oxygen. It may cause an explosion or fire.
Never use the equipment in the hyperbaric oxygen therapy chamber.
It may cause an explosion or fire.
Never operate the equipment where flammable gas or fluid such as
anesthetic, oxygen, and hydrogen are used.
It may cause an explosion or fire.
Содержание LX-8000 Series
Страница 1: ......
Страница 2: ......
Страница 22: ...xx ...
Страница 26: ...1 General Description 2 ...
Страница 34: ...3 Preparation 10 ...
Страница 40: ...5 Respiration Monitoring 16 ...
Страница 54: ...6 SpO2 Monitoring 30 ...
Страница 74: ...9 Other Setting Items 50 ...
Страница 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Страница 84: ...11 Troubleshooting 60 ...
Страница 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Страница 102: ...14 Standard and Optional Accessories 78 ...
Страница 108: ...15 Specification 84 ...
Страница 109: ......
Страница 110: ......