4. ECG Monitoring
12
Two leads such as lead I and II can be measured. Lead III, aVR, aVL, and aVF
can be also displayed from the setting on the receiving monitor. For details,
refer to the operation manual of the receiving monitor
For 5-electrode (chest) lead cable
Limb Lead
+Chest
One limb lead and one chest lead (brown) measurements are available.
Standard Limb leads can be selected from lead I, lead II, or lead III.
Refer to “8. Operation”.
The chest lead waveform is measured from the chest lead (brown) positioned
on the chest.
■
Attaching the Electrodes
CAUTION
Always use the same type of electrodes. If different types of electrodes
are used at the same time, the difference between the polarization
potential from each electrode may interfere with monitoring.
Do not reuse the disposable electrodes. It is intended for single patient
use only.
Clean the electrode sites with alcohol wipes
or other skin preparation. If necessary,
shave the electrode sites to remove
excessive hair.
Peel off the disposable electrode
Pay attention not to touch the electrode gel.
Attach the lead cable end to the electrode
(convex part).
Turn right and left to verify that it is securely
attached.
Brown
(V)
Red
(LL)
Green
(RL)
White
(RA)
Black
(LA)
Содержание LX-8000 Series
Страница 1: ......
Страница 2: ......
Страница 22: ...xx ...
Страница 26: ...1 General Description 2 ...
Страница 34: ...3 Preparation 10 ...
Страница 40: ...5 Respiration Monitoring 16 ...
Страница 54: ...6 SpO2 Monitoring 30 ...
Страница 74: ...9 Other Setting Items 50 ...
Страница 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Страница 84: ...11 Troubleshooting 60 ...
Страница 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Страница 102: ...14 Standard and Optional Accessories 78 ...
Страница 108: ...15 Specification 84 ...
Страница 109: ......
Страница 110: ......