5
3. Preparation
1) Installing the Batteries
The LX-8300M functions with two “AA” size (“LR6” size) alkaline batteries.
The battery operation time of LX-8300M is as follows.
When
SpO
2
measurement is ON: Approximately 2.5 days (new batteries)
Conditions: When measuring ECG, RESP, SpO
2
with default settings,
operating temperature 23
C
When
SpO
2
measurement is OFF: Approximately 6.5 days (new batteries)
Conditions: When measuring ECG, RESP with default settings, SpO
2
measurement OFF, operating temperature 23
C
*Disconnecting the SpO
2
probe does not satisfy the above condition.
It is necessary to set the SpO
2
measurement to OFF. Refer to “Turning
OFF the SpO2 measurement“.
However, continuous operating time may be shorter than the above mentioned
time depending on the application of the SpO
2
probe (sensor).
WARNING
Unplug the ECG lead cable when the battery compartment lid is opened.
Otherwise, patient leakage current beyond the allowable value may
occur.
Use new "AA" size (“LR06” size) alkaline batteries.
Do not short out the (+) and (-) terminals. It may result in exothermic
heat and fire, the leakage from the batteries may damage the
equipment, or an explosion from the batteries may occur.
Install the batteries with the correct polarity.
Do not charge alkaline batteries. Any attempt to charge the batteries
may cause them to leak or break.
Do not use a disassembled or a damaged battery due to drop or shock.
The leakage from the batteries may damage the equipment, or an
explosion from the batteries may occur.
Do not use different types of batteries at the same time. The leakage
from the batteries may damage the equipment, or an explosion from the
batteries may occur.
Remove the exhausted batteries immediately. The leakage from the
batteries may damage the equipment, or an explosion from the batteries
may occur.
If the equipment is not in use for a long period of time, remove the
batteries and store the equipment in an appropriate place. If the
batteries are left in the equipment for a long period of time, the leakage
from the batteries may damage the equipment or an explosion from the
batteries may occur.
Make sure to replace the two batteries simultaneously. If a new and
used battery are mixed, a leakage from the batteries may damage the
equipment or an explosion from the batteries may occur.
Содержание LX-8000 Series
Страница 1: ......
Страница 2: ......
Страница 22: ...xx ...
Страница 26: ...1 General Description 2 ...
Страница 34: ...3 Preparation 10 ...
Страница 40: ...5 Respiration Monitoring 16 ...
Страница 54: ...6 SpO2 Monitoring 30 ...
Страница 74: ...9 Other Setting Items 50 ...
Страница 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Страница 84: ...11 Troubleshooting 60 ...
Страница 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Страница 102: ...14 Standard and Optional Accessories 78 ...
Страница 108: ...15 Specification 84 ...
Страница 109: ......
Страница 110: ......