15. Specification
83
Safety
General Standard:
ANSI / AAMI ES 60601-1: 2005(R)2012 and
A1:2012, C1:2009(r)2012 and A2: 2010(r)2012
(Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance)
EMC Standard:
IEC 60601-1-2: 2014
(Medical electrical equipment – Part 1-2: General
requirements for basic safety and essential
performance – Collateral standard:
Electromagnetic disturbances – Requirements and
tests)
The class of protection
against electric shock:
Internally Powered Equipment
The type of protection
against electric shock:
ECG/RESP: Type CF Applied Part
SpO
2
: Type CF Applied Part
Operating Environment
Temperature:
10
C to 40
C / 50
F to 104
F
Humidity:
30% RH to 85% RH (non-condensing)
Atmospheric Pressure:
70 kPa to 106 kPa
Transport / Storage Environment
Temperature:
-10
C to 60
C / 14
F to 140
F
Humidity:
10% RH to 95% RH
(40
C / 104
F, non-condensing)
Atmospheric Pressure:
70 kPa to 106 kPa
Содержание LX-8000 Series
Страница 1: ......
Страница 2: ......
Страница 22: ...xx ...
Страница 26: ...1 General Description 2 ...
Страница 34: ...3 Preparation 10 ...
Страница 40: ...5 Respiration Monitoring 16 ...
Страница 54: ...6 SpO2 Monitoring 30 ...
Страница 74: ...9 Other Setting Items 50 ...
Страница 76: ...10 Changing the Transmitter Channel and Group ID 52 ...
Страница 84: ...11 Troubleshooting 60 ...
Страница 94: ...13 Maintenance and Inspection 70 Periodic Check List ...
Страница 102: ...14 Standard and Optional Accessories 78 ...
Страница 108: ...15 Specification 84 ...
Страница 109: ......
Страница 110: ......