Printer Adjust
1) Printer Status
Monitors the magazine open/close status, recording paper status, thermal head temperature
information.
・
Magazine
Status
Set Nil
Magazine is closed.
Magazin is opened.
・
Recording paper
Status
Set Nil
Paper is present.
Paper is out.
・
Thermal Head Temperature
The temperature display is
1℃
resolution.
Status
Normal Abnormal
Thermal Head temperature is below
70
℃
.
When the thermal head
temperature is 70
℃
or above, it will
be detected as abnormal. Once this
condition is detected, printing and
feeding will stop until the
temperature drops to 60
℃
or
below.