Inspection of Forceps
5.9
CAUTION
Do not bend forceps with a curvature radius of 10 mm or less.
Bending forceps with a small curvature may break it.
(1) Inspect the operation of the forceps.
Visually check the forceps for breakage or significant
bends, and for sharp edges on protrusions that may injure
the patient.
(2) Form the spring of the forceps into a ring approximately
200 mm in diameter.
(3) Operate the handle of the forceps and check that their tips
open and close.
200mm
Chapter 5 Preparation for Use of Ultrasonic Endoscope
5-26
EB530US_E2_202B1237458F.indb 26
2013/12/23 11:51:02