Attachment and Inspection of Balloon
5.7
WARNING
Use a disinfected and sterilized balloon installation tool.
There is a risk of infection.
Do not use a balloon on patients allergic to latex.
It may cause an anaphylactic reaction.
(1) Feed sterile water from the balloon water feed inlet using
a syringe and purge air from the balloon tubes.
[Note]
Feed water to the balloon manually by using a syringe.
Do not use any other instrument than syringe.
[Note]
The syringe can be attached to the balloon water feed
channel via an extension tube.
[Note]
If the syringe is not connected firmly, air may enter the
balloon.
[Note]
When using a screw syringe, do not tighten the screw
more than necessary.
(2) Prepare the balloon and balloon installation tool.
[Note]
Do not use any balloon after its expiration date has passed.
Handle 2
Handle 1
Chapter 5 Preparation for Use of Ultrasonic Endoscope
5-21