(7) Connect the US connector to the Ultrasonic Endoscope
connector of the ultrasonic processor.
Turn the lock handle of the US connector clockwise to
(8)
lock the connector.
<Combination with the EPX-2500 system>
(1) Insert the LG connector of the Ultrasonic Endoscope into
the Endoscope socket on the processor.
Insert the EVE connector of the Ultrasonic Endoscope into
(2)
the 500 system connector socket on the processor.
(3) Attach the suction button.
Use the level of disinfection of sterilization suitable to
the application before using the forceps valve and suction
button.
Set the direction of the suction connector at the
“ ” position of the index, and insert the arm into the
Endoscope.
Rotate the suction connector in the direction of the arrow.
To remove the suction connector, return it to the index
position, and pull out the entire equipment.
Chapter 5 Preparation for Use of Ultrasonic Endoscope
5-9
EB530US_E2_202B1237458F.indb 9
2013/12/23 11:50:57