Inspecting Suction, Water Feed and Forceps Channel
5.5.3
WARNING
Firmly attach the forceps valve to the forceps inlet of the Ultrasonic
Endoscope.
There is a risk of infection.
(1) Switch on the power to the suction unit, cart and processor
and light source.
Keep the lamp off.
Have a glass of sterile water ready.
(2)
(3) Put a forceps valve in forceps inlet.
Dip the distal end of the Ultrasonic Endoscope in water,
and check that depressing the suction button sucks in
water and that releasing it stops suction.
[Note]
Check that the forceps valve is correctly connected to the
forceps inlet.
[Note]
Do not use any expired forceps valve.
In the case of
EPX-4400 System
In the case of EPX-2500 System
Chapter 5 Preparation for Use of Ultrasonic Endoscope
5-13