Inspection of Distal End of Ultrasonic Endoscope, US Connector and Cable
5.5.4
(1) Turn the power on by pressing the standby switch on the
ultrasonic processor or on the keyboard.
Turn the power on by pressing the STANDBY switch on
the front side of the ultrasonic processor or STANDBY
key on the left hand side of the keyboard.
(2) Check to confirm by touching with a finger that the
transducer of the distal end of Ultrasonic Endoscope is not
hot.
(3) Visually check the distal end of Ultrasonic Endoscope,
the US connector and the cables for abnormalities such as
cracks or dents etc.
In case of the SU-7000
Chapter 5 Preparation for Use of Ultrasonic Endoscope
5-15
EB530US_E2_202B1237458F.indb 15
2013/12/23 11:50:59
Содержание EB-530US