5-34
006-240-50 2006.10
Troubleshooting
5.6 Failing in Auto F.D.C. (Automatic Density Calibration)
If fails in automatic density calibration processing for any reason, an error window like that
shown below appears.
[Fig. 5.44]
Press the
button to perform automatic density calibration processing again.
If the same failure recurs after Auto F.D.C. processing is performed again, a film will be
printed by using a density table before the failure has occurred. If the density is normal,
continuous use is possible. If the error occurs again, report it to the nearest service center.
If fails in automatic density calibration processing after a film pack has been supplied, an
error window like that shown below will open.
[Fig. 5.45]
Press the
button to chose it, and then Use Utility operation mode the perform
automatic density calibration processing.
Содержание Drypix 5000
Страница 7: ...1 1 006 240 20 2004 03 1 Chapter 1 Introduction ...
Страница 17: ...2 1 006 240 20 2004 03 2 Chapter 2 System Overview ...
Страница 23: ...3 1 006 240 20 2004 03 3 Chapter 3 Basic Operation ...
Страница 40: ...3 18 006 240 20 2004 03 Basic Operation ...
Страница 41: ...4 1 006 240 20 2004 03 4 Chapter 4 Utility Operation ...
Страница 87: ...5 1 006 240 20 2004 03 5 Chapter 5 Troubleshooting ...
Страница 125: ...6 1 006 240 20 2004 03 6 Chapter 6 Regular Care and Maintenance ...
Страница 130: ...6 6 006 240 60 2007 03 Regular Care and Maintenance ...
Страница 131: ...7 1 006 240 20 2004 03 7 Chapter 7 Specifications ...
Страница 136: ...7 6 006 240 20 2004 03 Specifications ...
Страница 138: ...Maintenance and Inspection 006 240 20 2004 03 ...
Страница 141: ......
Страница 142: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A ...