Utility Operation
1
2
3
4
5
6
7
006-240-40 2006.09
4-5
4
At the press of the
button, the automatic film density calibration will be performed.
Selecting the
icon on
[Fig. 4.6] Tray Select Screen
will enable you to finely adjust a
maximum density (Dmax) value. You can use this function when you wish to change the maximum
density with respect to the setup value, depending on the view box luminance or room brightness.
To perform automatic film density calibration by selecting a shift value for high-density calibration
(Dmax), select a desired tray on the
[Fig. 4.6] Tray Select Screen
and then
press
.
[Fig. 4.7] Shift Value (Dmax) Select Screen
2.5
3.6
Shift value=2
Shift value=1
Shift value=0
Shift value=–1
Shift value=–2
REFERENCE
Density
Image input value
If the shift value (Dmax) is changed, densities over
OD=2.5 will change.
Changing the shift value by 1 step will change the
density by about 0.1 at OD=3.6.
At the press of the
button here with a
shift value for high-density calibration (Dmax)
selected accordingly, the shift value applicable to
automatic film density calibration will be deter-
mined.
If the
button is pressed, the display
will return to the
[Fig. 4.6] Tray Select Screen
.
If the
button is pressed on the
[Fig. 4.6] Tray Select Screen
, automatic film
density calibration processing will be performed.
The range of variation of the density shift value by automatic film density calibration is presented
to be served as a guideline.
The specification of this equipment provides that the maximum density is 3.6 and a density that
exceeds 3.6 will not be guaranteed.
Note, however, that the maximum density specified for the DI-ML film is 4.0.
Содержание Drypix 5000
Страница 7: ...1 1 006 240 20 2004 03 1 Chapter 1 Introduction ...
Страница 17: ...2 1 006 240 20 2004 03 2 Chapter 2 System Overview ...
Страница 23: ...3 1 006 240 20 2004 03 3 Chapter 3 Basic Operation ...
Страница 40: ...3 18 006 240 20 2004 03 Basic Operation ...
Страница 41: ...4 1 006 240 20 2004 03 4 Chapter 4 Utility Operation ...
Страница 87: ...5 1 006 240 20 2004 03 5 Chapter 5 Troubleshooting ...
Страница 125: ...6 1 006 240 20 2004 03 6 Chapter 6 Regular Care and Maintenance ...
Страница 130: ...6 6 006 240 60 2007 03 Regular Care and Maintenance ...
Страница 131: ...7 1 006 240 20 2004 03 7 Chapter 7 Specifications ...
Страница 136: ...7 6 006 240 20 2004 03 Specifications ...
Страница 138: ...Maintenance and Inspection 006 240 20 2004 03 ...
Страница 141: ......
Страница 142: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A ...