4.10 Automated F.D.C.
Use this function to set the timing for achieving automatic film density calibration when supplying
a new film pack.
The operation procedure is as follows.
1
Start up the Utility. (Refer to section 4.2.1.)
2
Press the
button on the Utility initial screen (2/3).
[Fig. 4.59]
3
A screen for selection will appear. Select any for
When Execute?
and press the
button. The quit window will then open. Press the
button.
[Fig. 4.60]
•
By each film pack:
Performs automatic film density calibration every time when a film pack is supplied.
•
By each film lot No.:
Performs automatic film density calibration every time when a film lot number is changed.
A slight difference may result in the density due to differences in sensitivity among film lots.
Содержание Drypix 5000
Страница 7: ...1 1 006 240 20 2004 03 1 Chapter 1 Introduction ...
Страница 17: ...2 1 006 240 20 2004 03 2 Chapter 2 System Overview ...
Страница 23: ...3 1 006 240 20 2004 03 3 Chapter 3 Basic Operation ...
Страница 40: ...3 18 006 240 20 2004 03 Basic Operation ...
Страница 41: ...4 1 006 240 20 2004 03 4 Chapter 4 Utility Operation ...
Страница 87: ...5 1 006 240 20 2004 03 5 Chapter 5 Troubleshooting ...
Страница 125: ...6 1 006 240 20 2004 03 6 Chapter 6 Regular Care and Maintenance ...
Страница 130: ...6 6 006 240 60 2007 03 Regular Care and Maintenance ...
Страница 131: ...7 1 006 240 20 2004 03 7 Chapter 7 Specifications ...
Страница 136: ...7 6 006 240 20 2004 03 Specifications ...
Страница 138: ...Maintenance and Inspection 006 240 20 2004 03 ...
Страница 141: ......
Страница 142: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A ...