Regular Care and Maintenance
5
Use a gauze moistened with alcohol or ethanol anhydride to wipe off water remaining
on the roller.
The cleaning roller will be dried in a few minutes.
6
Put the roller back to its original position and close the front cover.
Do not raise the lever with hand, otherwise a problem such as damages to gears or film
stuck may occur with the equipment.
[Fig. 6.4]
Lever
Cleaning roller lever operation
caution label
If a cleaning roller that is not dried enough is used, film density may not be output correctly
and unevenness may occur on the film surface.