55
Section 2: RealEar evaluation with FM system
2.3.7 Test aid with high-level input—RHA90
1. Set the client and hearing aid up for real-ear testing as described in Section 2.3.1.
Make sure to level the sound field speaker.
2. From the SPL Test Screen, press [MENU].
3. Select PT_BURST for AIDED CURVE 1, AIDED CURVE 2, and AIDED CURVE 3.
4. Press [MENU] to return to the SPL Test Screen.
5. Press
[
∧
] to select CRV 1.
6. Use [>] to adjust the source level to 90 dB SPL.
7. Press [START/STOP] to take measurement.
8. This curve will be referred to as the Real-Ear Hearing Aid Saturation Response or
RHA90. See Figure 2.3.7.
Figure 2.3.7
RHA90-Real-Ear Saturation Response of the Aid