36
Testing FM Systems with FONIX 6500-CX Analyzers
1.6.4 Test aid with typical input — EHA65
1.
Set up the hearing aid for testing in the sound field by following the directions
found in the hearing aid setup in Section 1.4.1. If you’ve already leveled the sound
field speaker, you can skip to step 6 in that section.
2.
Press [MENU] on the remote module.
3.
Press [
∨]
to highlight CREATE TARGET.
4.
Press [<, >] to select SPL TARGET.
5.
Press [START/STOP] twice. This will open the SPL Test Screen.
6. Look in the bottom left corner of your screen. If it says AUTO, press [UNAIDED] on
the remote module. This will put you in MANUAL Mode.
7. Press [MENU] to enter the SPL Menu.
8. Select COMP for AIDED CURVE 1, AIDED CURVE 2, and AIDED CURVE 3.
9. Press [MENU] to return to the SPL Test Screen.
10. If necessary, select CRV 1 with [
∧
].
11. Use [<, >] to select a source level of 65 dB SPL.
Figure 1.6.4
Hearing aid response at 65 dB SPL