41
Section 1: Coupler Tests of FM Systems
1.6.9 Test environmental mic response with high-level input — EEF90
1. Set up the environmental microphone for testing as described in Section 1.4.3. Set
the FM receiver to FM+ENV. If the sound field speaker is already leveled, you don’t
have to re-level it.
2. Press
[
∨
] to select CRV 3.
3. Press [START/STOP] to take the measurement.
4. This measurement will be referred to as EEF90 or the Electroacoustic FM and
Environmental Response with 90 dB SPL input to the environmental microphone.
5. Compare EEF90 (CRV 3) to EFE90 (CRV 2) and EHA90 (CRV 1). All three curves
should match. See Figure 1.6.9.
Figure 1.6.9
Comparison of EHA90, EFM90, and EEF90, the environmental
microphone’s response at 90 dB SPL