31
Section 1: Coupler Tests of FM Systems
7. Press [MENU] to enter the SPL Menu.
8. Select COMP for AIDED CURVE 1 and AIDED CURVE 2.
9. Press [MENU] to return to the SPL Test Screen.
10. If necessary, select CRV 1 with [
∧
].
11. Use [<, >] to select a source level of 65 dB SPL.
12. Press [START/STOP] to begin the measurement.
13. Press [START/STOP] once the measurement has stabilized. This curve is the
Electroacoustic Hearing Aid Response or EHA65. See Figure 1.5.1.
1.5.2 Test FM microphone with typical input — EFM80
1. Set up the FM system for testing according to FM setup described in Section 1.4.2.
2. Use
[
∨
] to select CRV 2.
3
Use [<, >] to set the RMS source to 80 dB SPL.
4.
Press [START/STOP] to start the measurement.
5. Once the measurement stabilizes, press [START/STOP] again to stop the measure-
ment.
Figure 1.5.2
Comparison of EHA65 (CRV1), the hearing aid response at 65 dB
SPL and EFM80 (CRV2), the FM response at 80 dB SPL