NOTICE!
Fritsch mills are speed controlled. The devices are
equipped for this with frequency converters. In order to
comply with the EMC directive, many measures must
be taken to prevent operational transient emissions.
The possible leakage currents resulting from filtering
measures can trigger a conventional residual current
circuit breaker in the mains line. This is no defect!
To prevent this, special residual current circuit breakers,
which are adapted for operation with frequency con-
verters, are commercially available.
Operation without a residual current switch is possible,
but must be done in accordance with the relevant regu-
lations.
Installation
- 23 -