2.8.4 Imbalance detection
The device switches off if there is excessive imbalance. (see
6.6 „Mass balance“ on page 39 and
linktarget [Unwuchtschalter] doesn't exist but
@y.link.required='true'
)
2.8.5 Operation on GFCI (Ground-Fault Circuit Interrupters)
It is possible that the leakage currents will become marginal during oper-
ation. This may trigger the earth leakage circuit breaker. These values can
quickly be reached when all devices are added to the circuit.
Solution: Circuit without an earth leakage circuit breaker or, if possible,
increase the earth leakage circuit breaker threshold.
Safety information and use