8
Cleaning
DANGER!
Mains voltage!
– Before beginning with cleaning work, disconnect
the mains plug and protect the device against being
unintentionally switched back on!
– Do not allow any liquids to flow into the device.
– Indicate cleaning work with warning signs.
– Put safety equipment back into operation after
cleaning work.
When cleaning the entire device, adhere to the guide-
lines of the Accident Prevention Regulation (BGV A3) -
especially when the device has been set up in a dusty
environment or when the grinding stock processed pro-
duces dust.
8.1 Grinding chamber
To access the grinding chamber, the fixed crushing jaw (12) must first be
removed.
1.
Pull out the source material container (8).
2.
Open the lid (7).
3.
Set the gap width to the biggest distance.
4.
Turn the guide pin (2) in such a way that its groove is above the
position screw, which provides access to the guide pin (2).
5.
Then, pull the guide pin (2) out of its guide.
Cleaning
- 43 -
Содержание PULVERISETTE 1 premium Series
Страница 17: ...3 8 Protection class IP 44 Technical data 17 ...
Страница 54: ...14 Safety logbook Date Maintenance Repair Name Signature Safety logbook 54 ...
Страница 56: ...V Voltage 16 W Warning information 10 WEEE 49 Weight 16 Index 56 ...
Страница 57: ......