24
Installation
1
Connection, measuring unit
2
Outlet
3
Connection for external pump
4
Inlet for liquids (max. 0.5 bar)
5
Connection for measuring unit (only if the ultrasonic box is used)
6
Mains connection socket (only if the ultrasonic box is used)
1.
Connect the supplied data cable to the "Measuring Unit" connection (1) of
the dispersion unit and to one of the "Dispersion Units" connections of the
measuring unit.
2.
Connect the supplied outlet hose with inner diameter of 25 mm to the
outlet connector (2) and secure with the hose clamp included.
3.
Connect the wet dispersion unit with your building
connection; use the pressure regulator included. The
pressure regulator is equipped with an additional flow
limiter and thus reduces the pressure to 0.5 bar!
The "Inlet for liquids" connection (4) is used to fill the system with max. 0.5
bar. Connection the hose for the liquid to your building connection or liquid
storage tank and to the connection (4) on the dispersion unit. Secure the
hose with the hose clamps included.