39
Using the device
5.
In this project case you can select the SOP "F16 Standard measure" by
clicking on "New measurement".
SOP refers to a Standard Operating Procedure used in
our software. In an SOP, work steps are pre-defined and
processed with a single keystroke.
Ensure that you select the SOP which is suitable for your
device.
6.
The measuring window with the live image, camera and flash settings
opens.
7.
If necessary and depending on the setting in the SOP, the intensity of the
feeder can vary. Once the beam absorption is correct, you can start the
measurement.
The current beam absorption is displayed in the lower
line as ‘good, high and low density’.
8.
Once the cancellation criteria "Image count" or "Particle count" are
reached, the measurement is finished and you can display the evaluation
by clicking on "Stop" and "Close.
9.
When the measurement is finished, remove the sample from the collecting
vessel and clean the parts that come into contact with the samples before
you begin to measure other materials.
5.5
Fritsch test powder F70
You can check the precision and accuracy of your device
by measuring the Fritsch test powder at least once a week,
depending on how often you use the device. This will indicate
contamination or malfunctions in your system.
Measure the supplied Fritsch test powder F70 with the ANALYSETTE 28 Image-
Sizer as follows:
1.
Set the device to I (On) with the main switch.
2.
Start the Image Sizing software.
3.
Create a new folder in the archive.
4.
A "New Project Case" must be created In the newly created folder.