70
Name and Status
As explained previously, the unit name (or Comment) is
displayed on the first row and the second row indicates the
logger status.
If the logger is running, then the following screen is displayed.
Mark Unit on PC
To indicate the logger icon in the software’s Map View, press
the
Enter
button when this screen is displayed. The logger’s
icon located in the Map View will blink a number of times.
Batt Level
(battery level)
If the power adapter is not connected the battery percentage
will be displayed.
If the power adapter is connected
External power
will be
displayed.
Signal Level Test
The Signal Level test allows the user to check the signal
strength and quality of the link between the logger and its
parent unit. This test is used to determine optimal positioning
of the logger during deployment.
Batt Level:
External power
Batt Level: 67%
Fourier Systems
Logger running
Signal level:
Run test -> Menu