157
6. If you selected
All
inputs, then by default the logger will
be setup to log data on
Input 1
during the calibration
process. Otherwise, if you selected a specific input, it will
setup the logger to calibrate on that input.
Logger Data Pane
The Logger Data pane displays real-time readings on the
logger, at a default sampling rate of one per second. You can
also view the logger’s general status. Use the Logger Data
pane to verify your logger is properly calibrated.
The Calibration dialog will enable either Two-point or Offset
calibration, or both, depending on the sensor selected.
Note:
The calibrated sensor parameters will be saved both in
the DataNet logger and in the software memory, so there is no
need to calibrate the logger
every time you run the software.
4.14.3.
Performing a Two-point
Calibration
To perform a two-point calibration, you must have two
reference points to input against the real logger sensor
values.
Figure 52:
Two-point calibration window
1. In
the
Point #1
field, enter the first
Reference Value
and
the corresponding
Logger Value
.