154
Sensor
Type
Calibration
Method
Current
4 – 20 mA
Two-point
calibration
Humidity (Internal) Digital
Two-point
calibration
Temperature
(Internal)
Digital
Two-point
calibration
Temperature
(Internal)
PT-100
Two-point
calibration and
offset
Temperature PT-100
2-wire
Two-point
calibration and
offset
Temperature
Thermocouple J
Offset calibration
Temperature
Thermocouple K
Offset calibration
Temperature
Thermocouple T
Offset calibration
Voltage
0 to 1 V
Two-point
calibration
Voltage
0 to 50 mV
Two-point
calibration
Calibration Tips
Prior to any two-point calibration it is recommended to restore
factory calibration defaults.
Prior to calibrating any of the thermocouple sensor types,
calibrate the Voltage 50 mV sensor as this will set the gain
(slope) of all thermocouple sensor types.
After calibrating 50 mV, TC-J, TC-K or TC-T sensor types for
all inputs you can refine the offset calibration for each input
individually.