151
7. You can also click
Check All
to select all the sensor
alarms for notification.
Note:
If the logger transmission rate is slower than the
sampling rate, an alarm notification is still sent even if the
logger transmission time hasn’t passed. For example, if
the sampling rate is every 5 minutes, and the
transmission rate is every 10 minutes, if a logger reaches
a high alarm after 7 minutes a notification will be sent.
The system won’t wait until the scheduled transmission
time.
4.13.4.
Email and SMS Notification
Formats
Email Notification Format
When the Email notification is sent the recipient receives the
Email in the following format:
Email Header:
DataNet Alarm Internal Temperature (Last recorded 25.47C)
Low Alarm
Email Body:
DataNet Alarm
Comment: Factory 1/F
S/N: 808932
08-07-09 13:53:41
Internal Temperature Low Alarm
Last recorded 25.47 C
The Email header contains the sensor type, last recorded
sample, and alarm type.
Содержание DataNet