Jan 31st, 2005
- 91 -
Material Emissivity
1.0µm
1.6µm
8-14µm
“Gob”
n.r.
0.9
n.r.
Gravel n.r.
0.95
0.95
Gypsum n.r.
0.4-0.97
0.8-0.95
Ice n.r.
—
0.98
Limestone n.r.
0.4-0.98
Paint (non-Al.)
—
0.9-0.95
0.9-0.95
Paper (any color)
n.r.
0.95
0.95
Plastic (opaque, over 20 mils)
n.r.
0.95
0.95
Rubber n.r.
0.9
0.95
Sand n.r.
0.9
0.9
Snow n.r.
—
0.9
Soil n.r.
—
0.9-0.98
Water n.r.
—
0.93
Wood, Natural
n.r.
0.9-0.95
To optimize surface temperature measurement accuracy consider the following:
1. Determine the object emissivity for the spectral range of the instrument to be used for
the measurement.
2. Avoid reflections by shielding object from surrounding high temperature sources.
3. For higher temperature objects use shorter wavelength instruments, whenever possible.
4. For semi-transparent materials such as plastic film and glass, assure that the background
is uniform and lower in temperature than the object.
5. Hold instrument perpendicular to surface whenever emissivity is less than 0.9. In all
cases, do not exceed angles more than 30 degrees from incidence.
Содержание Ti30
Страница 4: ...Appendix B Infrared Theory Q A 92 Glossary 96 Jan 31st 2005 4 ...
Страница 25: ...Man Machine Interface Flow Charts Main Loop Jan 31st 2005 25 ...
Страница 26: ...Image Recording Procedure Emissivity Adjustment Procedure Jan 31st 2005 26 ...
Страница 27: ...Reflected Temperature Compensation Adjustment Procedure Recall Images Procedure Jan 31st 2005 27 ...
Страница 38: ...Figure 30 Choose Destination Location Figure 31 Configuring and Installing InsideIR Jan 31st 2005 38 ...
Страница 70: ...Jan 31st 2005 70 ...
Страница 86: ...Jan 31st 2005 86 ...