Tip:
When focusing an image, look for identifiable patterns or shapes on the object
surface that can help establishing the best image definition. Objects contours,
dividing lines or limits are particularly useful.
Proper focusing is not only
important to the image quality, but also greatly affects the accuracy of
temperature measurements.
One way to verify if the unit is properly focused is
to find what focus adjustment produces the highest temperature indication on the
display (keep E and RTC values constant, preferably E set at 1.0 and RTC disabled).
Selecting the Color Palette
The selected color scheme is mostly a matter of personal preference. In some cases, it makes
sense to use a specific palette for specific applications. The rainbow palette provides more
contrast between areas with close temperature values; on the other hand the ironbow palette
may provide more visual comfort in some cases because the colors blend in smoothly.
However, despite the popularity of color palettes, the gray scale is recommended for most
measurements because it is easier for the human eye to discern subtle thermal changes on
gray tones than colors.
Tip:
Always begin with the gray scale palette while you get a feel for the thermal scene
you are viewing. Then, work with the available measurement modes and the Level
and Gain adjustments before choosing which color palette to use.
Selecting the Measurement Mode
Measurement modes are simply different ways to present thermal information on the
display. Depending on the mode you choose, you may display all temperature points shown
on the thermal image, or you may select a narrow interval of temperature points being
displayed at one time. The decision of which mode to use depends on your needs and
limitations. From a thermal resolution standpoint, you are better off working with a narrow
interval of temperature points, because you will be able to see very subtle temperature
differences, since you have more colors or gray tones to represent fewer temperature points;
on the other hand, if you are just looking for larger temperature differences, you can not
work with a narrow temperature interval because important temperature values might not be
displayed.
There are three measurement mode positions associated with this switch:
Automatic Semi
Automatic Manual
In
Automatic
mode the Ti30 imager automatically adjusts the image to show the
lowest temperature value present in the thermal scene (MIN) and the highest temperature
Jan 31st, 2005
- 58 -
Содержание Ti30
Страница 4: ...Appendix B Infrared Theory Q A 92 Glossary 96 Jan 31st 2005 4 ...
Страница 25: ...Man Machine Interface Flow Charts Main Loop Jan 31st 2005 25 ...
Страница 26: ...Image Recording Procedure Emissivity Adjustment Procedure Jan 31st 2005 26 ...
Страница 27: ...Reflected Temperature Compensation Adjustment Procedure Recall Images Procedure Jan 31st 2005 27 ...
Страница 38: ...Figure 30 Choose Destination Location Figure 31 Configuring and Installing InsideIR Jan 31st 2005 38 ...
Страница 70: ...Jan 31st 2005 70 ...
Страница 86: ...Jan 31st 2005 86 ...