Autoset Options
On delivery or after a reset, the Autoset function captures
waveforms
≥
15 Hz and sets the input coupling to DC.
Note
Setting the Auto Set adjust to 1 Hz slows down
the Auto Set response. The display shows
LF-AUTO.
To configure Auto Set for capturing slow waveforms down
to 1 Hz:
1.
Push
to open the Menu.
2.
Use
YZ
to highlight
USER OPTIONS
.
3.
Push
to open the USER OPTIONS menu.
4.
Use
YZ
to highlight
Autoset Settings
.
5.
Push
to open the USER > AUTOSET menu.
6.
Use
YZ
to highlight
Search For Signals >1 Hz
.
7.
Push
to accept the change and exit the menu.
To configure Autoset to maintain the actual input coupling
(AC or DC), continue from step 5 above:
6.
Use
YZ
to highlight
Couplings Unchanged
.
7.
Push
to accept the change and exit the menu.
Grounding Guidelines
XW
Warning
To prevent possible electrical shock, fire, or
personal injury, use only one COM (common)
connection
˜
or ensure that all connections
to COM
˜
are at the same potential.
Incorrect grounding can cause problems. Use these
guidelines for proper grounding:
•
Use the short ground leads when measuring DC or
AC signals on input A and input B. See Figure 8, item
4 on page 17.
•
Use the unshielded black ground lead to COM
(common) for Ohm (
Ω
), Continuity, Diode, and
Capacitance measurements. See Figure 7, item 1 on
page 16.
•
Use of the unshielded ground lead is also possible for
single or dual input measurements for waveforms with
a frequency up to 1 MHz. This may add some hum or
noise to the waveform display due to the unshielded
ground lead.
68
Содержание ScopeMeter 123B
Страница 8: ...123B 124B 125B U ivatelsk p ru ka vi...
Страница 10: ...123B 124B 125B U ivatelsk p ru ka viii...
Страница 104: ...123B 124B 125B Users Manual vi...
Страница 106: ...123B 124B 125B Users Manual viii...
Страница 123: ...4 Industrial ScopeMeter Scope and Meter Mode A B 5 A B hxv04 eps Figure 8 Correct Grounding Setup 17...
Страница 124: ...6 7 123B 124B 125B Users Manual A B 10 mV A 1 mV A hxv57 eps Figure 9 Temperature and Current Measurement Setup 18...