Industrial ScopeMeter®
Scope and Meter Mode
Waveform Acquisition
The Test Tool can be setup to acquire waveforms as
needed for a specific application. This section explains
the setup options.
Single Acquisition
To catch single events, you can perform a single shot
(one time screen update).
To set up the Test Tool for a single shot on the input A
waveform:
1.
Connect the probe to the signal to be measured.
2.
Push
to open the Scope and Meter button bar.
3.
Push
2
to open the SCOPE SETTINGS menu.
4.
Use
YZ
to highlight the Update as
Single
.
5.
Push
to make the change.
6.
Push
4
to exit the menu.
The Test Tool will now show the acquisition screen
and the Information area updates:
Waiting
Test Tool is waiting for a trigger
Run
single acquisition is triggered
Hold
single acquisition is complete
hxv19.eps
To continue with the next single acquisition:
7.
Push
and wait for another single acquisition
trigger.
27
Содержание ScopeMeter 123B
Страница 8: ...123B 124B 125B U ivatelsk p ru ka vi...
Страница 10: ...123B 124B 125B U ivatelsk p ru ka viii...
Страница 104: ...123B 124B 125B Users Manual vi...
Страница 106: ...123B 124B 125B Users Manual viii...
Страница 123: ...4 Industrial ScopeMeter Scope and Meter Mode A B 5 A B hxv04 eps Figure 8 Correct Grounding Setup 17...
Страница 124: ...6 7 123B 124B 125B Users Manual A B 10 mV A 1 mV A hxv57 eps Figure 9 Temperature and Current Measurement Setup 18...