A
B
BB120
1
hxv53.eps
Figure 14. 10:1 Scope Probes
6.
Use
YZ
to highlight
Probe Adjust
.
Industrial ScopeMeter®
Maintenance
7.
Push
to open the PROBE ADJUST menu.
A square wave appears on the screen.
8.
Adjust the trimmer screw
in the probe housing to
give an optimum square wave.
9.
Push
4
to exit the menu.
Calibration Information
The Test Tool specifications are based on a 1 year
calibration cycle. Recalibration must be done by qualified
personnel. Contact your local Fluke representative for
more information about recalibration.
To find the firmware version and calibration date of your
Test Tool:
1.
Push
to open the Menu.
2.
Use
YZ
to highlight
USER OPTIONS
.
3.
Push
to open the USER OPTIONS menu.
4.
Use
YZ
to highlight
Information
.
63