123B/124B/125B
Users Manual
Noise Reduction
To view the waveform without higher frequency noise:
1.
Push
to open the Scope and Meter button bar.
2.
Push
3
to open the INPUT SETTINGS menu.
3.
Use
YZ
to highlight the Noise filter as
ON
.
4.
Push
to make the change.
This noise filter is a 10 kHz bandwidth limit filter.
hxv15.eps
Glitch Display
By default, the displayed waveform shows glitches. For
each time position the minimum and maximum values
since the last time position are shown. A glitch of 25 ns or
wider is still visible on the screen even when using a
slower timebase.
To disable this feature and see one sample/channel for
each sample time:
1.
Push
to open the Scope and Meter button bar.
2.
Push
2
to open the SCOPE menu.
3.
Use
YZ
to highlight the Type as Glitch Off.
This feature will be turned off for both channel A and
channel B.
4.
Push
to make the change.
5.
Push
4
to exit the menu.
24
Содержание ScopeMeter 123B
Страница 8: ...123B 124B 125B U ivatelsk p ru ka vi...
Страница 10: ...123B 124B 125B U ivatelsk p ru ka viii...
Страница 104: ...123B 124B 125B Users Manual vi...
Страница 106: ...123B 124B 125B Users Manual viii...
Страница 123: ...4 Industrial ScopeMeter Scope and Meter Mode A B 5 A B hxv04 eps Figure 8 Correct Grounding Setup 17...
Страница 124: ...6 7 123B 124B 125B Users Manual A B 10 mV A 1 mV A hxv57 eps Figure 9 Temperature and Current Measurement Setup 18...