123B/124B/125B
Users Manual
Rise Time Measurements
To measure rise time:
1.
Push
3
to select CURSOR ON.
2.
Push
1
to select
(rise time).
Observe that two
horizontal cursors are displayed.
3.
Push
4
if only one trace is displayed and select
MANUAL or AUTO. AUTO automatically does steps
5 to 7. For two traces, select the required trace A
or B.
4.
Use
YZ
to move the upper cursor to 100% of the
trace height. A marker is shown at 90%.
5.
Push
2
to select the other cursor.
6.
Use
YZ
to move the lower cursor to 0% of the
trace height. A marker is shown at 10%.
The reading now shows the risetime from 10%-90%
of the trace amplitude and the voltage at the cursors
in relation to the zero icon (-).
7.
Push
3
to disable the cursors.
hxv25.eps
34