Industrial ScopeMeter®
Recorder Mode
5.
Use
YZWX
and
to enter the time in hours
and minutes.
Events are used to determine how often the
measured reading deviates from the initial reading
when recording starts. It is easy to see the time of a
deviation when you view the recording after it stops.
6.
Use
YZ
to highlight
Set Threshold
.
7.
Push
and use
YZ
and
to enter the
percentage deviation for meter readings.
8.
Use
YZ
to highlight the memory type for recording
as either the internal memory of the Test Tool or an
SD memory card.
9.
Push
to accept the memory location.
10. Push
4
when done.
11. To start or stop a recording, push
or
4
.
The Test Tool continuously logs all readings to memory
and displays these as graphs. When both input A and
input B are on, the upper graph is input A.
Note
The Test Tool beeps when an event occurs. If
no event is specified, a beep happens when a
new minimum or maximum value is detected.
hxv40.eps
The Recorder shows a graph derived from the MAIN
readings.
The other readings show the average (AVG), the
minimum (MIN), and the maximum (MAX) reading since
the Recorder start and the time of the most recent change
of a value.
51
Содержание ScopeMeter 123B
Страница 8: ...123B 124B 125B U ivatelsk p ru ka vi...
Страница 10: ...123B 124B 125B U ivatelsk p ru ka viii...
Страница 104: ...123B 124B 125B Users Manual vi...
Страница 106: ...123B 124B 125B Users Manual viii...
Страница 123: ...4 Industrial ScopeMeter Scope and Meter Mode A B 5 A B hxv04 eps Figure 8 Correct Grounding Setup 17...
Страница 124: ...6 7 123B 124B 125B Users Manual A B 10 mV A 1 mV A hxv57 eps Figure 9 Temperature and Current Measurement Setup 18...