RUSKA 7050, 7050i & 7050LP
Users Manual
4-6
Menu | Setup
Setup is used to configure the system. This function includes setting all of the limits, user
parameters, user-defined units of measure, remote interface and system setup.
4-8.bmp
Figure 4-8. Menu | Setup - Menu
Menu | Setup — Limits
The Menu | Setup | Limits menu is used to setup all of the limits in the system. The
pressure limits can be used to protect the device-under-test (DUT) from overpressure.
glc17.bmp
Figure 4-9. Menu | Setup | Limits - Menu
High Limit
The High Limit — User Defined Maximum Pressure Limit is often set just over the
full-scale pressure of the device under test (DUT) in order to protect the DUT. The PPI
will then prevent the pressure from exceeding the high limit. If the user enters a value
greater than the user defined limit, the unit will not accept the value and it will generate
an error code informing the user that they have entered a set-point that is greater than the
high pressure limit. In control mode, if for any reason the pressure exceeds the
user-defined limit, it will change into the measure mode and again display an error
message to the operator informing them that the error occurred.
Содержание RUSKA 7050
Страница 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Страница 4: ......
Страница 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Страница 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Страница 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Страница 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Страница 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...
Страница 80: ...RUSKA 7050 7050i 7050LP Users Manual 7 4...